Exhaust air is double-filtered through high-quality ULPA filters with typical efficiency of 99.999% at 0.1~0.2um particles.
An angled cabinet front ensures an ergonomic working posture
Cabinet operates at negative pressure relative to the laboratory in order to prevent migration of pathogenic materials out of the work area.
UV lamp ensures work area sterilization effectively
Class III Biosafety Cabinet is totally enclosed and gas-tight with ULPA filtered supply and exhaust air. Work is performed with long-sleeved gloves. The cabinet is kept under negative pressure of at least 120Pa, and airflow is maintained by a dedicated exterior exhaust system. It can protect the operator, product and environment. It is designed for work with level 4 pathogens and provide an alternative to the positive-pressure suit made of maximum containment laboratories
Model | SKSSBC395B |
External Size(W*D*H,mm) | 1340*850*2100 |
Internal Size(W*D*H,mm) | 1240*650*650 |
Pass Box Size(W*D*H,mm) | Internal: 400*390*340 |
External: 480*400*400 | |
Gloves | One pair, 800mm butyl rubber gloves |
Front Window | 8mm toughened glass, anti ultraviolet radiation |
Display | LCD Display |
Pre-filter | Polyester fiber, washable |
ULPA Filter | Air supply filter: 99.999% efficiency at 0.1-0.2um |
First exhaust filter: 99.999% efficiency at 0.1-0.2um | |
Second exhaust filter: 99.999% efficiency at 0.1-0.2um | |
UV Lamp | 30W*1, 8W*1 |
Fluorescent lamp | 28W*2 |
Noise | ?62dB |
Airflow Volume | 440~870cfm |
Main Material | 304 stainless steel |
Pressure | -120Pa |
Alarm | Visual and audio alarm |
Illumination | ?800Lux |
Power Supply | 110/220V±10%, 50/60Hz |
Consumption | 700W |
Gross Weight | 395KG |
Package Size(W*D*H) | 2000*1000*2220mm |
Standard Accessory | Fluorescent lamp, UV Lamp*4, Gloves, Pressure meter, Drain valve, Remote control |